| Judul Artikel | ANALISIS KELUARAN BERKAS FOTON 10 MV DENGAN VARIASI SOURCE TO SURFACE DISTANCE DAN FIELD SIZE PADA PESAWAT TERAPI LINAC (STUDI KASUS : RSUD PROF. DR. MARGONO SOEKARJO PURWOKERTO) |
| Abstrak (Bhs. Indonesia) | Pesawat terapi Linear Accelerator (LINAC) digunakan dalam radioterapi untuk menghasilkan berkas foton berenergi tinggi sehingga ketepatan keluaran dosis menjadi faktor penting dalam keberhasilan terapi dan keselamatan pasien. Penelitian ini bertujuan untuk menganalisis keluaran dosis berkas foton 10 MV pada variasi Source to Surface Distance (SSD) dan Field Size berdasarkan protokol IAEA TRS-398 serta mengevaluasi kesesuaiannya terhadap batas toleransi ±3%. Pengukuran dilakukan menggunakan fantom air dan detektor ionisasi Farmer chamber pada variasi SSD 90 cm, 95 cm, dan 100 cm serta field size 5×5 cm², 10×10 cm², dan 15×15 cm². Hasil penelitian menunjukkan bahwa peningkatan SSD menyebabkan keluaran dosis menurun, sedangkan peningkatan field size menyebabkan keluaran dosis meningkat akibat bertambahnya radiasi hambur. Dari sembilan kondisi pengukuran, hanya dua kondisi yang memenuhi batas toleransi ±3%, yaitu SSD 95 cm dengan field size 5 × 5 cm² dengan deviasi 0,50% dan SSD 100 cm dengan field size 10 × 10 cm² dengan deviasi 0,40%, sedangkan tujuh kondisi lainnya melebihi batas toleransi dan memerlukan adjustment. Hasil penelitian menunjukkan bahwa variasi SSD dan field size berpengaruh signifikan terhadap keluaran dosis berkas foton 10 MV pada pesawat LINAC dan perlu diperhatikan dalam proses quality assurance dan quality control (QA/QC) radioterapi. |
| Abtrak (Bhs. Inggris) | The Linear Accelerator (LINAC) is used in radiotherapy to produce high energy photon beams, making accurate dose output an important factor in treatment efficacy and patient safety. This study aimed to analyze the dose output of a 10 MV photon beam under variations in Source to Surface Distance (SSD) and Field Size based on the IAEA TRS-398 protocol and to evaluate its compliance with the ±3% tolerance limit. Measurements were performed using a water phantom and a Farmer chamber ionization detector at SSDs of 90 cm, 95 cm, and 100 cm, with field sizes of 5 × 5 cm², 10 × 10 cm², and 15 × 15 cm². The results showed that increasing the SSD decreased the dose output, while increasing the field size increased the dose output due to increased scattered radiation. Of the nine measurement conditions, only two met the ±3% tolerance limit, namely an SSD of 95 cm with a field size of 5 × 5 cm², with a deviation of 0.50%, and an SSD of 100 cm with a field size of 10 × 10 cm², with a deviation of 0.40%. The other seven conditions exceeded the tolerance limit and required adjustment. The results indicate that variations in SSD and field size significantly affect the dose output of the 10 MV photon beam from the LINAC and should therefore be considered in radiotherapy quality assurance and quality control (QA/QC) processes. |
| Kata kunci | LINAC, foton 10 MV, keluaran dosis, SSD, field size, IAEA TRS-398. |
| Pembimbing 1 | Efita Pratiwi Adi, S.Pd., M.Sc |
| Pembimbing 2 | Dr. Kartika Sari, S.Si., M.Si |
| Pembimbing 3 | |
| Tahun | 2026 |
| Jumlah Halaman | 61 |
| Tgl. Entri | 2026-08-19 01:34:39.540191 |
|---|