Judul ArtikelPEMODELAN DAN ANALISIS KARAKTERISTIK
PERCENTAGE DEPTH DOSE DAN PROFIL DOSIS PADA
PESAWAT LINAC 6 MV DENGAN VARIASI LUAS
LAPANGAN PENYINARAN
Abstrak (Bhs. Indonesia)Pemeriksaan kendali mutu pesawat Linear Accelerator (Linac) harus mencakup
pemeriksaan berkas radiasi. Simulasi merupakan salah satu alternatif dalam
pengukuran berkas radiasi ketika proses pengukuran langsung pada pasien sulit
dilakukan karena banyaknya tahapan proses. Parameter yang menunjukkan
kualitas berkas radiasi ditunjukkan oleh Percentage Depth Dose (PDD) dan profil
dosis. Penelitian ini bertujuan untuk menganalisis karakteristik percentage depth
dose dan profil dosis pada pesawat Linac 6 MV dengan variasi luas lapangan
penyinaran. Penelitian dilakukan dengan menggunakan metode Monte Carlo
program PHITS. Program PHITS digunakan untuk memodelkan kepala pesawat
Linac Elekta Precise sebagai sumber radiasi dan phantom air sebagai objek
penyinaran. Luas lapangan penyinaran yang digunakan bervariasi yaitu berukuran
5 5, 10 10, 15 15, 20 20 dan 25 25 cm2
. Pemodelan pesawat Linac dan
phantom air telah berhasil. Modelnya divisualisasikan dalam gambar 2D dan 3D.
Hasil simulasi menunjukkan jejak partikel melewati komponen-komponen kepala
pesawat Linac hingga menembus phantom air. Kurva PDD di daerah build-up
menunjukkan bahwa pada lima variasi luas lapangan penyinaran diperoleh dosis
maksimum berada pada kedalaman 1.5 cm. Kurva profil dosis pada lima variasi
luas lapangan penyinaran dan lima kedalaman menghasilkan nilai flatness yang
berbeda-beda. Nilai flatness yang diperoleh pada lima variasi luas lapangan
penyinaran di kedalaman 10 cm berkisar antara 0.70%-1.95%. Homogenitas profil
dosis di kedalaman 10 cm menunjukkan hasil yang baik dengan nilai flatness <3%
sesuai batas toleransi klinis.
Abtrak (Bhs. Inggris)Quality assurance procedures for linear accelerators (Linas) must include
comprehensive evaluation of the radiation beam. In situations where direct
measurement on patients is impractical due to the complexity of clinical
procedures, simulation serves as a viable alternative. Key indicators of beam
quality include the Percentage Depth Dose (PDD) and dose profile. This study
aims to investigate the characteristics of PDD and dose profiles for a 6 MV Linac
across various field sizes using the Monte Carlo method implemented through the
PHITS simulation program. The Elekta Precise Linac head was modeled as the
radiation source, with an air phantom used as the irradiation medium.
Simulations were conducted for five square field sizes: 5×5, 10×10, 15×15,
20×20, and 25×25 cm². The Linac head and air phantom models were
successfully constructed and visualized in both 2D and 3D. The simulation results
demonstrated particle trajectories traversing the Linac head components and
penetrating the air phantom. PDD curves in the build-up region showed that the
maximum dose for all field size variations occurred at a depth of 1.5 cm. Dose
profile analysis at various depths and field sizes yielded distinct flatness values.
At a depth of 10 cm, the flatness ranged between 0.70% and 1.95%. Overall, dose
profile homogeneity at this depth was found to be clinically acceptable, with
flatness values remaining below the 3% tolerance threshold recommended in
clinical guidelines.