VALIDASI METODE ANALISIS KETOKONAZOL PADA SAMPEL URIN MANUSIA MENGGUNAKAN EKSTRAKSI FASA PADAT DAN KROMATOGRAFI CAIR KINERJA TINGGI
Abstrak (Bhs. Indonesia)
Validasi metode penentuan ketokonazol dalam urin manusia menggunakan ekstraksi fasa padat dan kromatografi cair kinerja tinggi (KCKT) telah dilakukan. Kondisi optimum KCKT diperoleh pada laju alir 1 mL/menit, komposisi fase gerak yang mengandung metanol:air (85:15), deteksi UV-Vis pada panjang gelombang 220 nm dan waktu analisis 3 menit. Nilai koefisien determinasi (R2) yang diperoleh sebesar 0.9998, batas deteksi (LOD) sebesar 0,340 ppm dan batas kuantifikasi (LOQ) sebesar 1,133 ppm, standar deviasi (SD) sebesar 0,192, nilai KV sebesar 1,296 % dan nilai selektivitas sebesar 3,01. Metode ekstraksi fasa padat menggunakan oktadesilsilane (C18) sebagai adsorben telah dioptimasi. Optimasi ekstraksi fasa padat dicapai menggunakan eluen metanol dan volume sampel sebesar 1 mL. Rata-rata % perolehan kembali (% recovery) ketokonazol dalam sampel urin sebesar 87,50 %, dengan simpangan baku relatif (RSD) 1,76 (n=3). Hasil analisis ekstraksi fasa padat-KCKT untuk ketokonazol dalam sampel urin yang di spike diperoleh sebesar 4,375 ppm.
Abtrak (Bhs. Inggris)
Method validation of ketoconazole determination in human urine sample using solid phase extraction and high performance liquid chromatography (HPLC) has been performed. HPLC optimum conditions was obtained at a flow rate of 1 mL / min, the composition of the mobile phase containing methanol:water v/v (85:15), UV-Vis detection at a wavelength of 220 nm and a 3-minute analysis time. The coefficient of correlations (r) was obtained of 0,99992, the limit of detection (LOD) of 0,340 ppm and the limit of quantification (LOQ) of 1,133 ppm, standard deviation (SD) of 0,192, KV value amounting to 1,296% and the selectivity of 3,01. Solid phase extraction method using oktadesilsilane (C18) as adsorbent has been optimized. Optimization of solid phase extraction is achieved using methanol eluent and sample volume of 1 mL. Average % recovery of ketoconazole in the urine sample of 87,50%, with a relative standard deviation (RSD) of 1,76% (n = 3). The results of ketoconazole analysis in the spike urine sample by SPE-HPLC was 4,375 ppm.
Kata kunci
Pembimbing 1
Pembimbing 2
Pembimbing 3
Tahun
Jumlah Halaman
Save